The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
History

Sat, 01 Mar 2025 17:30:00 +0000

Type Values Removed Values Added
Description Contec Health CMS8000 Patient Monitor sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device. The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Wed, 12 Feb 2025 21:15:00 +0000

Type Values Removed Values Added
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'total'}, 'version': '2.0.3'}


Fri, 31 Jan 2025 17:15:00 +0000

Type Values Removed Values Added
Description The affected product sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device. Contec Health CMS8000 Patient Monitor sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.
Metrics cvssV3_1

{'score': 7.5, 'vector': 'CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H'}


Fri, 31 Jan 2025 16:45:00 +0000


Thu, 30 Jan 2025 18:30:00 +0000

Type Values Removed Values Added
Description The affected product sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.
Title Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Weaknesses CWE-912
References
Metrics cvssV4_0

{'score': 7.7, 'vector': 'CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N'}


cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published: 2025-01-30T18:17:29.717Z

Updated: 2025-03-01T17:12:24.132Z

Reserved: 2025-01-21T17:29:48.728Z

Link: CVE-2025-0626

cve-icon Vulnrichment

Updated: 2025-01-31T15:37:36.861Z

cve-icon NVD

Status : Awaiting Analysis

Published: 2025-01-30T19:15:14.010

Modified: 2025-03-01T18:15:34.140

Link: CVE-2025-0626